<rp id="pogfq"></rp>

    亚洲综合图片区自拍区 Skip to content Skip to navigation

    产品介绍

    Specifications:

      Wxx02 (2000 N (450lbs) Version) Wxx04 (4000 N (900 lbs) Version)
    Voltage
    Input voltage [VDC] 12 24 12 24
    Load
    Static load (Fx), maximum [N (lbs)] 2000 (450) 4000 (900)
    Dynamic load (Fx), maximum [N (lbs)] 2000 (450) 4000 (900)
    Stroke
    Stroke lengths, standard [cm (inch)] 10 (3.94), 20 (7.87), 30 (11.81), 40 (15.75), 50 (19.69)
    Speed
    Speed version 58A 54A 58A 54A 58A 58A
    Speed, no load [mm/s (inch/s)] 5.8 (0.23) 11.0 (0.43) 5.8 (0.23) 11.0 (0.43) 5.8 (0.23) 5.8 (0.23)
    Speed, rated load [mm/s (inch/s)] 4.0 (0.16) 8.0 (0.31) 4.0 (0.16) 8.0 (0.31) 4.0 (0.16) 4.0 (0.16)
    Current
    Current draw @ rated max. load [A] 4.5 9.0 2.2 4.5 9.0 4.5
    Current draw @ stall/in-rush [A] 14.0 21.0 8.0 11.0 21.0 11.0
    General Data
    End play, maximum [mm (inch)] 0.5 (0.020)
    Operating temperature limits [°C (°F)] -25 to +40 (-13 to +104) (1)
    Duty cycle, maximum [%] 10
    On-time, maximum [s] 180 90 180 90 90 90
    Life, average [cycles] 10,000
    Sound level [dBa] < 45
    Lead screw type acme
    Protection class IP67
    Anti rotation mechanism(2) yes
    Analog feedback output signal voltage [VDC] 0.5 - 4.5
    Analog feedback output signal linearity [%] 0.5
    Digital feedback input voltage [VDC] 3.8 - 24
    Digital feedback output resolution [pulses/mm] 10 5 10 5 10 10
    Compliances IEC 60601-1, CSA C22.2, AAMI ES60601-1 (3)
    Options
    • Electronic limit switches (ELS) (4)
    • ELS + low level switching (1) (5)
    • Analog or digital position feedback, with or without ELS
    • Mounting adapters turned 90°

    ?           1) Thrust is reduced at lower temperatures for 2000 N (450 lbs) high speed (54A) and 4000 N (900 lbs) units with ELS + low level switching (option E).
               2) Prevents the extension tube from rotating if it is not fixed in the end.
               3) 4 kN units greater than 30 cm of stroke are not ETL recognized. ETL certification available on N, E, P and D feedback options only.
               4) Shuts off power at the end of stroke and all along the stroke at overload conditions. ELS is normally set for 120% of the rated dynamic maximum load.
               5) Low level switching allows low level signals to control the actuator movement instead of having to switch the high amperage motor voltage.

    back to top